Simulation of imaging with Nal(TI) crystals and position-sensitive photomultiplier tubes
Document Type
Conference Proceeding
Publication Date
1-1-1994
Abstract
The imaging characteristics of miniature gamma cameras that consist of a single sodium iodide (NaI(Tl)) crystal coupled to a position-sensitive photomultiplier tube (PSPMT) have been studied via Monte Carlo simulations. Images obtained with such cameras using conventional position calculations exhibit considerable distortions, particularly compression. This study demonstrates that the observed distortions result from the non-uniform sensitivities of PSPMTs. Simulation results are compared to images obtained using a prototype single-crystal/single-PSPMT miniature gamma camera.
Publication Source (Journal or Book title)
IEEE Nuclear Science Symposium & Medical Imaging Conference
First Page
1358
Last Page
1361
Recommended Citation
Ordonez, C., Mintzer, R., Aarsvold, J., Yasillo, N., & Matthews, K. (1994). Simulation of imaging with Nal(TI) crystals and position-sensitive photomultiplier tubes. IEEE Nuclear Science Symposium & Medical Imaging Conference (pt 2), 1358-1361. Retrieved from https://repository.lsu.edu/physics_astronomy_pubs/3402