Purpose: To validate the performance of a respiratory gating system for the automated delivery of the deep inspiration breath-hold (DIBH) technique.
Methods: The gating system utilized an automatic gating interface (Elekta Response) which connected a marker-based respiratory motion monitoring system to the linear accelerator control system. The gating system was characterized dosimetrically and temporally using two distinct approaches. Central-axis output and energy constancy were evaluated across 8 beam-matched linear accelerators. Additionally, a representative set of 5 treatment plans were delivered both non-gated and gated to a 2D diode array (MapCHECK). The respiratory motion monitoring system optically tracked a reflective marker that was attached to a dynamic phantom (QUASAR). The phantom was programmed to replicate a typical DIBH breathing waveform. The passing rates between these modes of operation were evaluated using gamma analysis and a percent dose difference comparison. Modular and end-to-end approaches were used to quantify system latencies. The modular components evaluated were the streaming latency of the tracking camera, sampling rate of the tracking software, signal travel time, and latency of the linear accelerator. The end-to-end approach involved measuring the displacement of a target moving at known velocities during the during the gating process.
Results: Output and energy constancy were both within ± 0.5% for each beam energy and linear accelerator investigated. The average differences in passing rates between non-gated and gated modes of operation were within ± 0.4% using gamma analysis (2%, 1mm). Average passing rates between modes of operation were greater than 99% using a percent dose difference comparison (1%). The first gated segment was found to have significantly (p =.02) longer beam-on latency compared to the subsequent gated segment. End-to-end beam-on and beam-off latency for the subsequent gated segment was found to be 1.49 and 0.34 seconds, respectively, which was consistent with measured component totals.
Conclusion: The gating system was able to achieve dosimetric operating characteristics that are desirable for accurate delivery of the DIBH technique. The methodology presented can be generalized to other respiratory gating systems that utilize the automatic gating interface studied in this work.
Recommended Citation
Stock, Michael G., "Validation of a Respiratory Gating System for Automated Delivery of the Deep Inspiration Breath-hold Technique" (2021). LSU Master's Theses. 5341.
https://repository.lsu.edu/gradschool_theses/5341
Committee Chair
Fontenot, Jonas D.
DOI
10.31390/gradschool_theses.5341
Included in
Analytical, Diagnostic and Therapeutic Techniques and Equipment Commons, Medical Sciences Commons, Physics Commons