Title
Experimental studies of electroosmotic flow dynamics during sample stacking for capillary electrophoresis
Document Type
Article
Publication Date
7-15-2003
Abstract
Electroosmotic flow dynamics during a field-amplified sample stacking experiment have been studied experimentally using the periodic photobleaching of a dilute, neutral fluorophore added to the separation buffer. The effects of hydrodynamically injecting different sample plug lengths containing a mixture of arsenic compounds dissolved in 0.125 mM (120, 240, and 600 s) and 41.7 μM (27, 45, and 74 s) phosphate buffer with a separation buffer concentration of 12.5 mM phosphate buffer were examined. Changes in electroosmotic flow during sample stacking and separation were monitored at a rate of 1 Hz. The observed effects of increasing the sample plug length on electroosmotic flow and electrophoretic current agreed qualitatively with predictions by theoretical models presented in the literature. Electroosmotic flow changes on the order of 100% (1.6-3.3 mm/s) were observed. Broadening of the flow monitoring peaks has been used to examine parabolic flow due to the discontinuous buffer systems used for sample stacking.
Publication Source (Journal or Book title)
Analytical Chemistry
First Page
3531
Last Page
3538
Recommended Citation
Pittman, J., Gessner, H., Frederick, K., Raby, E., Batts, J., & Gilman, S. (2003). Experimental studies of electroosmotic flow dynamics during sample stacking for capillary electrophoresis. Analytical Chemistry, 75 (14), 3531-3538. https://doi.org/10.1021/ac034102u