Title
Design considerations for a two-dimensional rapid scanning fluorimeter
Document Type
Article
Publication Date
9-1-1979
Abstract
The design or a two-dimensional rapid scanning fluorimeter which eliminates much unneeded transfer optics is described. The advantages and disadvantages of commercially available optical systems are discussed with regard to preservation of image quality and relative cost. The combination of holographic polychromators and conventional lenses in the recommended design produces spectra as accurate as those obtained on most conventional fluorimeters. © 1979.
Publication Source (Journal or Book title)
Analytica Chimica Acta
First Page
361
Last Page
372
Recommended Citation
Warner, I., Fogarty, M., & Shelly, D. (1979). Design considerations for a two-dimensional rapid scanning fluorimeter. Analytica Chimica Acta, 109 (2), 361-372. https://doi.org/10.1016/S0003-2670(01)84265-3