Abstract

Objective: The purpose of this study was to compare body composition measurements estimated by multi-frequency bioelectrical impedance analysis (MF-BIA) with air displacement plethysmography (ADP) in individuals with obesity. Methods: Bariatric patients were recruited from Geisinger's Center for Nutrition and Weight Management Clinic in Danville, Pennsylvania. Sixty-two participants (age = 52.4 ± 9.3 years; body mass index = 38.9 ± 8.0 kg m−2) reported for a same-day testing visit. Body composition was measured using a common MF-BIA analyzer (InBody 720, Biospace Co., Beverly Hills, CA) and ADP. Results: Strong relationships were observed between MF-BIA and ADP methods (r = 0.88–0.96, P < 0.001). There were no differences between MF-BIA and ADP measures of per cent body fat, fat mass or fat-free mass for the total sample or when examined by gender. Conclusions: The InBody 720 MF-BIA analyzer produced body composition measurements that were similar to ADP supporting the use of this technology in the obese population.